| Catalog Number |
856135012 |
| Medical Device Problem Code |
Device Contaminated During Manufacture or Shipping (2969)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/05/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during incoming receiving inspection at the distributorship, the device was found to be non-conforming due to having a bug-like substance under the shrink package.No patient or customer was involved in the event.Attempts have been made and no further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).G2: foreign - event occurred in japan customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected: b4; b5; d2; d9; g1; g3; g6; h1; h2; h3; h6.Visual evaluation of the returned product found a bug sealed in the shrink film.Reported event did not occur in an operating room or as part of a medical procedure; medical records are not available for review.Review of the device history records identified no deviations or anomalies during manufacturing.The root cause of the reported event can be attributed to a manufacturing issue.The event was confirmed.If any further information which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|