| Model Number |
DRIVER SHAFT, T-15, MEDIUM |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/10/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
On 11/10/2025, it was reported by a sales representative via (b)(4) that an ar-9545-t15-02 driver shaft was broken.No further information was provided.This was discovered during the case with no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
-complaint allegation is not confirmed.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause is attributed to user error of the device due to over-torquing/over-engaging the driver within the screw head.
|
| |
|