Investigation of this event is pending and a supplemental report will be sent upon its completion.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
A supplemental report is being submitted for investigation completion.Product event summary: the battery (b)(6) was not returned for evaluation.Review of the manufacturing documentation confirmed that the associated device met all requirements for release.Log file analysis was not performed since log files were not available for analysis.As a result, the reported event could not be confirmed due to insufficient evidence.Applicable risk documentation and experience with events of similar circumstances were considered; possible root causes of the reported battery not charging event can be attributed, but not limited, to an internal battery communication error, faulty integrated circuit, improper weld, soldering defect, out of temperature range, and/or a charge time-out of eight (8) hours.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|