| Catalog Number |
EN8818 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/07/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The lot number was reviewed for complaint trend, nonconforming report and capa.Devices met specification prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed in veeva to be associated.The device was not returned for evaluation.Without the benefit of analyzing the device, coloplast cannot comment on the condition of the device.Should additional information prompt us to alter or supplement any information or conclusions contained in this report, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
According to the available information, the device was explanted & replaced due to a tubing leak leading from the cylinder entrance.
|
| |
|
Event or Problem Description
|
|
According to the available information, the device was explanted & replaced due to a tubing leak leading from the cylinder entrance.
|
| |
|
Additional Manufacturer Narrative
|
|
Titan pump, cylinders 1 and 2, and reservoir were received for evaluation.Abrasion was noted on the inlet tubing of the pump.No functional abnormalities were noted with the pump.A separation was noted on the exhaust tubing of cylinder2 near the strain relief.This was a site of leakage.The surfaces appear to be rough and irregular, indicating stress was exerted.No functional abnormalities were noted with cylinder 1.No functional abnormalities were noted with the reservoir.Based on examination of the returned product, it was concluded that the rough and irregular surfaces associated with the separation indicate that sufficient stress was exerted on the exhaust tubing of cylinder 2 to separate the site while in-vivo.A separation of this type could then allow the loss of fluid, making the device inoperable.A review of the device history record confirmed the devices from this lot met all specifications prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed to be associated.
|
| |
|