| Model Number |
P8040H |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/25/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The baxter technician found the casters needed to be replaced.Per the baxter service manual the bed should be subject to an effective maintenance program.An annual service of the bed is advised in order to maintain its characteristics and performance.Brake casters should be checked for cuts, wear and quality of tread, etc.And replaced when necessary.Apply he brake and make sure the stretcher does not move.If there is movement, look at the brake components for wear.Apply the steer, and make sure the stretcher steers correctly.Look at the steer components for wear.Put the stretcher in neutral.Make sure all four casters rotate and roll freely.Adjust or replace components if necessary.A search of the baxter maintenance records showed baxter performed preventive maintenance on this bed on nov 25, 2025.It is unknown if the facility performed any other preventive maintenance on this bed.The technician replaced the casters to resolve the reported event.Based on this information, no further action is required.
|
| |
|
Event or Problem Description
|
|
On 25-nov-2025, a company representative - service personnel contacted technical service to report that trauma stretcher frame (product code: p8040h000487, serial number: (b)(6), had brakes not holding.This occurred before use.There was no patient/user injury, medical intervention, symptom, or adverse event reported.
|
| |
|