MAUDE Adverse Event Report: SYNTHES GMBH AO REAMING ATTACHMENT FOR TRS; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH AO REAMING ATTACHMENT FOR TRS; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT Back to Search Results
Catalog Number 05.001.210
Medical Device Problem Code Excessive Heating (4030)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during service and evaluation, it was determined that the attachment device had heat, was making an unexpected noise, and there was foreign substance/debris/cleaning/sterilization.It was further determined that the device failed pretest for check general function in running mode and check for runout inspection in running mode.This event did not occur during surgery.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported; however, it was reported that the event occurred in 2025.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: the actual device was returned without an alleged deficiency.During routine service and repair, it was determined that the device was generating heat.Therefore, the reported condition was confirmed.The assignable root cause of this condition was determined to be traced to component failure due to wear.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AO REAMING ATTACHMENT FOR TRS
Common Device Name
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer Contact
elba bello
eimattstrasse 3
oberdorf 04436
SZ   04436
9083863534
MDR Report Key23687117
Report Number8030965-2025-11972
Device Sequence Number9584337
Product Code HWE
UDI-Device Identifier07611819977891
UDI-Public(01)07611819977891
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number05.001.210
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/11/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/19/2025
Initial Report FDA Received Date12/01/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/04/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-