| Model Number |
E163 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/27/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the health care professional (hcp) called technical services (ts) for further review of this implantable cardioverter defibrillator (icd) presenting electrogram (egm).Upon review, the presenting egm showed myopotential oversensing noise on the right atrial (ra) and right ventricular (rv) lead channels.Ts provided programming and lead revision recommendations.It was noted that the ra lead is a non-boston scientific lead.No adverse patient effects were reported.The icd and both leads remain in service.
|
| |
|