| Model Number |
H93811 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/05/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that eight (8) sterile repeater pump tube sets had particulate matter in an unspecified location.The particulate was described as dark particles and fiber.This was observed during setup.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, h3, h4, h6 (update codes), h11.H11: eight (8) unused devices were received for evaluation.Visual inspection was performed on all the samples which identified six (6) samples with tiny dark particles embedded on the blue cap white connector inside the packaging material and not in the fluid pathway, and two (2) samples with tiny dark particles embedded on the white spike port inside the packaging material and not in the fluid pathway.The particulates were located outside the fluid path and therefore would not be delivered in the final product or pose any risk of patient harm.The reported condition was verified.The cause of the condition could not be determined; however, possibly inherent to contamination during the manufacturing or packaging process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Search Alerts/Recalls
|