Catalog Number 150621003
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Pain (1994); Swelling/ Edema (4577)
Date of Event 10/28/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this refers to a revision that took place in an investigator initiated clinical study (attune cementless rsa study).One patient in this study recently underwent a full revision at 11 months post-op.The patient had swelling and pain upon presentation to the clinic in (b)(6) (unscheduled visit).An interim ct-rsa was performed which showed no discernible migration of their tibial tray, no obvious migration that would suggest loosening.The patient was revised on (b)(6) 2025 for infection (stage 1 of 2-stage procedure).Dr.Performed the revision and noted the 4 peripheral pegs of the tray were solidly ingrown.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: an analysis of the product could not be performed since a physical sample was not received for evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Based on the information available, there is no indication that a design or manufacturing issue has caused the reported complaint condition; therefore, it has been determined that no corrective and preventive action is required at this time.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product is received at a later date, the investigation will be updated as applicable.Recaptured codes on h6 health effect - clinical code.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.An evaluation of the manufacturing record could not be performed as the required lot number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: a device history record (dhr) review or manufacturing records evaluation (mre) was not possible because lot code provided is not a valid finished goods lot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE FB TIB BASE SZ 3 POR
Common Device Name
KNEE IMPLANT - TIBIAL TRAY
Manufacturer (Section D)
DEPUY IRELAND - 3015516266
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
CORK MFG & MATERIAL WAREHOUSE
loughbeg ringaskiddy
cork
EI  
Manufacturer Contact
elba bello
700 orthpaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key23688553
Report Number1818910-2025-20902
Device Sequence Number15832946
Product Code MBH
UDI-Device Identifier10603295507406
UDI-Public(01)10603295507406(17)330731(10)JA03A0500
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K213781
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number150621003
Device Lot NumberJA03A0500
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/10/2025
Supplement Date Received by Manufacturer12/13/2025
03/19/2026
Initial Report FDA Received Date12/01/2025
Supplement Report FDA Received Date12/15/2025
03/20/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE FEM POR CR LT SZ 4; ATUNE CR LT MS INS SZ 4 6
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-