| Model Number |
PXCAT0 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Blurred Vision (2137); Visual Impairment (2138)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A non-health care professional reported that following an intraocular lens (iol) implant procedure, the patient experienced blurry vision.Clinical reason being mentioned as multifocal iol intolerance.The iol was explanted and exchanged for an unknown monofocal lens in the secondary implant procedure.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected information provided in b.2., and b.5.Based on information received following submission of the initial report, this event does not meet criteria for reporting as a serious injury.No further reports required.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
Additional information received and reported that the iol was exchanged for the different lens model but half a diopter more power indicating the correct lens power was not implanted initially.There is no reported defect or fault with the iol.The patient's issues was resolved and there was no further impact to the patient.In surgeon¿s prognosis patient condition was good.
|
| |
|