MAUDE Adverse Event Report: TANDEM DIABETES CARE T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY; ALTERNATE CONTROLLER ENABLED INSULIN INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TANDEM DIABETES CARE T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY; ALTERNATE CONTROLLER ENABLED INSULIN INFUSION PUMP Back to Search Results
Model Number 1002717
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/22/2025
Type of Reportable Event Malfunction
Event or Problem Description
A cartridge alarm was reported by the caller during the load process, and it was confirmed that there was no adverse impact to the customer's blood glucose level.The issue occurred because the caller did not wait for the prompt to remove the used cartridge, leading to recurring alarms, despite efforts to load a new cartridge.As a corrective action, technical support educated the caller on correct cartridge loading procedures and decided to replace the pump.The caller will contact their healthcare provider for a backup insulin therapy, as none was available.Instructions on using storage mode, returning the faulty pump, and steps for the replacement pump were provided, and the caller acknowledged these instructions.A replacement pump was dispatched to the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY
Common Device Name
ALTERNATE CONTROLLER ENABLED INSULIN INFUSION PUMP
Manufacturer (Section D)
TANDEM DIABETES CARE
12400 high bluff drive
san diego CA 92130
Manufacturer Contact
michael trier
12400 high bluff drive
san diego, CA 92130
858 401-1451
MDR Report Key23692606
Report Number3013756811-2025-269563
Device Sequence Number18735709
Product Code QFG
UDI-Device Identifier00850006613380
UDI-Public00850006613380
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K201214
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number1002717
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/22/2025
Supplement Date Received by Manufacturer01/30/2026
Initial Report FDA Received Date12/02/2025
Supplement Report FDA Received Date02/02/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/18/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age76 YR
Patient SexFemale
Patient Weight100 KG
-
-