| Model Number |
DTMA1Q1 |
| Medical Device Problem Code |
Failure to Charge (1085)
|
| Health Effect - Clinical Codes |
Sepsis (2067); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/29/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Continuation of d10: 694765 lead implanted (b)(6) 2008 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that there was high pacing, defibrillation, and superior vena cava (svc) coil impedance, a lead integrity alert (lia) due to high rate non-sustained episodes and impedance variations, and a noise warning on the right ventricular (rv) lead.A charge circuit timeout occurred on the cardiac resynchronization therapy defibrillator (crt-d).The lead and device remain in use. no patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the device was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated charge circuit timeout.Analysis of the device memory showed the battery indicator signifying that it is time for device replacement.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was further reported that the crt-d was programmed off and the patient was fitted with a lifevest.Additionally, the patient expe rienced sepsis.The patient was transferred to a new hospital for an extraction of the entire crt-d system.No further patient complications have been reported as a result of this event.
|
| |
|