| Catalog Number |
110030777 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
11/03/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported the patient underwent a reverse shoulder revision procedure approximately 5 months post-implantation due to component dislocation.Attempts have been made, and no further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).G2: foreign - the event occurred in australia.Multiple mdr reports were filed for this event.Associated reports are available for the applicable concomitant products below in the d10 narrative.D10: concomitant medical products, part (lot): -00434906606(66623400) associated product information: -180559(62212850) -115398(66508387) -110032410(67109285) -180552(67160072) -180550(67221332) -180550(67232909) -00-4349-016-13(66866727) h6: proposed code: mechanical (g04)- head.The customer has not indicated whether the product will be zimmer biomet for investigation, and the product location is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found that would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent a shoulder revision approximately five months post-implantation due to dislocation with humeral polyethylene liner dissociation.Approximately seven weeks post-implantation, the patient lifted and carried weights contrary to postoperative instructions and sustained a shoulder dislocation.The patient presented to a local emergency department where closed reduction was attempted for several hours.The treating surgeon believed the prolonged reduction attempts resulted in humeral polyethylene liner disassociation, which led to the subsequent revision.It was reported that no further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated/corrected: b4; b5; b7; d2; d9; g1; g3; g6; h1; h2; h3; h10.The reported event cannot be confirmed.Visual examination of the provided photos identified the explanted products with signs of use and biological debris.However, as the products were not returned, further analysis could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Medical records were not provided.The root cause of the reported issue is attributed to patient non-compliance.It was indicated that the patient admitted to lifting and carrying weights in excess of the surgeons direct orders.Ifu indicates "premature failure of implants by loosening, fracture, dislocation, subluxation, and/or wear are more likely to occur where patients have unrealistic functional expectations (and conduct excessive, unusual, and/or awkward movement and/or activity), are overweight, or excessively overweight, or where patients fail to follow through with the required rehabilitation program.Attempts have been made to gather all product identification information and no further information has been provided.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|