• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMVIE US CORP LLC IMPL TWIST MP-1 5.0 MM 10 MM LZ; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMVIE US CORP LLC IMPL TWIST MP-1 5.0 MM 10 MM LZ; DENTAL IMPLANT Back to Search Results
Catalog Number 1994-LZ
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/14/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).G4: additional pma/510(k) number k962106.
 
Event or Problem Description
It was reported mount disengagement.Before use, the mount could not be removed because it had come loose from the fixture.Completed using a different fixture.No rough handling occurred within the clinic.Tooth site # 14.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
Zimvie did not receive one (1) 1994-lz, (impl twist mp-1 5.0 mm 10 mm lz) for evaluation.Visual inspection could not be performed.The investigation has been performed based on the available information using the item number, device history record (dhr) review, and risk management file (rmf).Dhr review was completed for the subject lot number 2023050697.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2023050697 for similar events and no other complaint was identified.The customer did not submit images for the reported event.Based on the investigation and risk management file review the most likely root cause determined from the investigation was improper screw placement combined with shipping vibrations.Therefore, based on the available information, a device malfunction was not verified.Without device receipt, the reported event is non-verifiable.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.The following sections have been updated: b4: date of this report.G3: date received by manufacturer.G6: checked "follow-up".H6: entered evaluation codes.H10: added manufacturer narrative.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPL TWIST MP-1 5.0 MM 10 MM LZ
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
mark  mashburn
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key23693752
Report Number0001038806-2025-03492
Device Sequence Number18714163
Product Code DZE
UDI-Device Identifier00889024013179
UDI-Public(01)00889024013179(17)280629(10)2023050697
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K013494
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number1994-LZ
Device Lot Number2023050697
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date12/02/2025
Supplement Report FDA Received Date01/27/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/29/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
Patient EthnicityNon Hispanic
-
-