| Catalog Number |
14200100 |
| Medical Device Problem Code |
Failure to Align (2522)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/11/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The reported event could not be confirmed during the investigation since the alleged failure was not reproduced during functional testing.The device inspection revealed the following: targeting arm was received and found damaged from its distal end.Apart from that the device shows normal signs of usage.A functional inspection was done on the returned targeting arm using a sample nail, tissue protection sleeve, drill sleeve and drill in which it was found that the drill is easily passing through the distal hole of the nail.Therefore, the device is functional, and the event is not confirmed.Given the device was confirmed to be fully functional even with slight damaged distal end, the root cause of the reported event could not be determined.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If more information is provided, the case will be reassessed.
|
| |
|
Event or Problem Description
|
|
The following information was reported: "several advanced gamma surgeons have since the introduction of gamma 4 experienced issues with the distal screw which misses the troch nail." this record covers surgery 1 out of 3.
|
| |
|
Event or Problem Description
|
|
The following information was reported: "several advanced gamma surgeons have since the introduction of gamma 4 experienced issues with the distal screw which misses the troch nail." this record covers surgery 1 out of 3.
|
| |
|
Additional Manufacturer Narrative
|
|
Once the investigation is completed, any additional information will be reported in a supplemental report.
|
| |
|