• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD INC. (COVINGTON) -1018233 BARDEX® LUBRICATH® HYDROGEL-COATED FOLEY CATHETER KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD INC. (COVINGTON) -1018233 BARDEX® LUBRICATH® HYDROGEL-COATED FOLEY CATHETER KIT Back to Search Results
Catalog Number 730116
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/20/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Initial reporter phone: (b)(6).Initial reporter fax: (b)(6).This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that after use, there were holes in the shaft of the foley catheter when there should be no holes, idc then had to be removed and replaced, doubling chance of infection and double up of procedure.
 
Additional Manufacturer Narrative
The reported event was inconclusive because this investigation did not result in any additional findings and no sample was available for evaluation.The reported event is addressed within the labeling.Warning: do not use ointments or lubricants having a petrolatum base.They will damage silicone.Visually inspect the product for any imperfections or surface deterioration prior to use.¿ use luer tip syringe to inflate.Note: aggressive traction, particularly in the presence of suturing, is not recommended for silicone foley catheters.To deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If necessary, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.No additional actions can be taken at this time.Correction: d.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that after use, there were holes in the shaft of the foley catheter when there should be no holes, idc then had to be removed and replaced, doubling chance of infection and double up of procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRICATH® HYDROGEL-COATED FOLEY CATHETER KIT
Common Device Name
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key23695544
Report Number1018233-2025-10662
Device Sequence Number16046228
Product Code EZC
UDI-Device Identifier00801741040672
UDI-Public(01)00801741040672(11)2220714(17)270429(10)NGGR2198
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K910846
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number730116
Device Lot NumberNGGR2198
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/20/2025
Supplement Date Received by Manufacturer01/22/2026
Initial Report FDA Received Date12/02/2025
Supplement Report FDA Received Date01/26/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/14/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-