Model Number 112266-001
Medical Device Problem Codes Difficult to Remove (1528); Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/16/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device was difficult to remove.The target lesion was located in the peripheral artery.A 2.4mm jetstream xc catheter was selected for a lower extremity angiogram procedure to treat peripheral artery disease (pad).During the procedure, resistance occurred when loading the non-boston scientific guidewire into the 2.4mm jetstream xc catheter.The 2.4mm jetstream xc catheter was difficult to advance and remove from guidewire inside of the patient.The catheter was ultimately removed from the guidewire by a pin and pull method.The same catheter was attempted for use again with a new guidewire, however there was resistance loading the catheter.The procedure was completed with an alternative device.There were no patient complications as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JETSTREAM? XC
Common Device Name
CATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
model farm road
cork T12 Y K88
EI   T12 YK88
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key23695617
Report Number2124215-2025-88406
Device Sequence Number15721347
Product Code MCW
UDI-Device Identifier08714729889922
UDI-Public(01)08714729889922(17)270822(10)37026228
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
K133023
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/02/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number112266-001
Device Catalogue Number112266-001
Device Lot Number0037026228
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/04/2025
Initial Report FDA Received Date12/02/2025
Date Device Manufactured07/22/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
.014 ABBOTT SPARTACORE GUIDEWIRE
Patient SexUnknown
-
-