BOSTON SCIENTIFIC CORPORATION EASYTRAK 3 IS-1; IMPLANTABLE LEAD Back to Search Results
Model Number 4549
Medical Device Problem Codes Over-Sensing (1438); Failure to Read Input Signal (1581)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/16/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the health care professional (hcp) wanted to review the presenting electrogram (egm) on this left ventricular (lv) lead.Technical services (ts) reviewed the device data, noted premature ventricular contraction (pvc) for some time and wondered if there is evidence of a possible dislodgment or an experienced fall.Additionally, alerts of low intrinsic amplitude out of range and pacing percentages below the programmed thresholds were noted.The plan is to continue to monitor and further evaluate in an in-clinic visit.No adverse patient effects were reported.This lv lead remains in service.Additional information was received that the lv lead showed deflections that could be potential oversensing.No adverse patient effects were reported.This lv lead remains in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYTRAK 3 IS-1
Common Device Name
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
CPI - DEL CARIBE
guidant puerto rico b. v.
#12, rd. #698
dorado PR 00646
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key23696773
Report Number2124215-2025-88348
Device Sequence Number21386429
Product Code LWP
UDI-Device Identifier00802526420399
UDI-Public(01)00802526420399(17)091108
Combination Product (Y/N)Y
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P010012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/02/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/08/2009
Device Model Number4549
Device Catalogue Number4549
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/26/2025
Initial Report FDA Received Date12/02/2025
Date Device Manufactured11/08/2007
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age58 YR
Patient SexMale
-
-