Catalog Number VCP493H
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/20/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent an unknown procedure on (b)(6) 2025 and suture was used.It was reported that the problem of pulling off suture needle happened in surgery.Total 8 products occurred needle pull off issue.Changed another one to complete the surgery.There were no patient consequences reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint#: (b)(4).Date sent to the fda: 12/2/2025.H6 component code: g07002 - no device problem found.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformance's were identified.Additional information was requested; the following was obtained: when were the needles detached/pulled off from the sutures (in the package, during removal from package, during handling prior to use on patient or during use on the patient)? please specify.Received information as following from sales rep via phone today.Please refer to the event description, other information requested is unknown.H3 investigational summary: the product was returned to eth for evaluation.Visual inspection revelated that twenty-one unopened foil packets were received for analysis.Product code vcp493h.Following the sampling plan, a visual inspection was conducted on 13 samples.No external damage was observed in the packets.The packets were open, and the swage and attachment area were noted to be as expected.The sutures were dispensed without issues.Although, the sutures in all the samples were noted with significant difference in suture¿s diameter and coloration at the tipping area.As per this condition observed in the sutures, the flex test was performed in the twenty-one samples.All sutures broke due to improper tipping, as the sutures were breaking away from the swage area.Based on the information currently available, damaged suture was identified during the investigation of the sample received.This product issue will be addressed through the ethicon quality system.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.Additional h3 investigational summary: the product was returned to eth for evaluation.Visual inspection revelated that one labeled winding former with one needle outside the foil packet was received for analysis.Product code vcp493.The returned needle was visually inspected, and the swage and attachment area were noted to be as expected.The barrel hole of the needle was examined under magnification, and suture remnant was observed.Additionally, the suture was not returned for evaluation, therefore no functional testing could be performed.A manufacturing record evaluation for the finished device batch number revealed no non-conformance's.Per the conditions of the returned sample, no conclusion could be reached regarding the cause of the reported complaint, and there was no information indicating patient involvement or adverse outcomes.As part of eth quality process all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCL+ UD 18IN 5-0 S/A P-3 PRM MP
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
3035526892
MDR Report Key23697159
Report Number2210968-2025-13482
Device Sequence Number21404623
Product Code GAM
UDI-Device Identifier10705031053663
UDI-Public(01)10705031053663(17)290930(10)103Z90
Combination Product (Y/N)Y
Initial Reporter CountryCH
PMA/510(K) Number
K032420
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/02/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberVCP493H
Device Lot Number103Z90
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/25/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/05/2025
Initial Report FDA Received Date12/02/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/10/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1