Product complaint # (b)(4).H3 investigation summary: the product was returned to ethicon for evaluation.One empty open foil and one labeled winding former with one detached were received for analysis.Product code vcp493h.During the visual inspection of the detached needle, the body and edge of the needle were noted with marks probably caused by a surgical instrument.This condition likely caused the needle to detach from the suture the barrel hole of the needle was examined under magnification and remnant suture was noted.In addition, the suture was not returned for evaluation.To avoid this kind of damage: grasp the needle in an area one-third (1/3) to one-half (1/2) of the distance from the attachment end to the point.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.The following information was requested, but unavailable: - when were the needles detached/pulled off from the sutures (in the package, during removal from package, during handling prior to use on patient or during use on the patient)? please specify --received information as following from sales rep via phone today.Please refer to the event description, other information requested is unknown.
|