| Model Number |
U1040245000 |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/03/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The following has been reported: on 11/3/2025 they received an ivenix complaint/disposable for evaluation and during evaluation they found a defect with a blonie product.See attached for description of issue.No adverse event reported no sample available pictures available findings: during sample evaluation, two defects were identified: -missing silicone barrier -cracked cap note: this was a k-zero connector on a fresenius kabi ivenix administration set.
|
| |
|
Search Alerts/Recalls
|