| Model Number |
VTICM5 12.1 (-10.00/2.5) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/01/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Manufacturer narrative: rotation, secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim# (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported that an implantable collamer lens was implanted into the patient's eye.The patient experienced lens rotation.The lens was repositioned, but then the lens was exchanged with a longer length lens and implanted vertically.The problem was resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: device evaluation: lens was returned in a microcentrifuge tube, dry, with residue/debris on the product.Visual inspection found the lens haptic broken.Dimensional inspection found the lens within specifications.Claim #: (b)(4).
|
| |
|