Corrected data: upon further review it was determined that section h4, device manufacture date in the initial mdr report, the date was incorrect.The correct date is aug 12, 2014.Therefore, this supplemental report is being submitted to provide the correct data.Additional narrative information: assessments of manufacturing, service, design, and usage were completed, and no product quality defects were found.Laser and power adjustment made.Attenuator and external power calibration completed.Room hvac was not working while testing on the first day.This appears to be an environmental issue.Reviews of labeling and risk management files were conducted, and no new issues/risks were identified.Johnson & johnson surgical vision will continue to monitor these types of complaints.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|