• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD INC. (COVINGTON) -1018233 BARDEX® LUBRICATH® TEMPERATURE SENSING FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD INC. (COVINGTON) -1018233 BARDEX® LUBRICATH® TEMPERATURE SENSING FOLEY CATHETER Back to Search Results
Catalog Number 119416M
Medical Device Problem Code Inaccurate Flow Rate (1249)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 11/21/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.Udi format updated with available product information per gudid.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that there had since been 3 more events with the similar situation where the foley catheter was not drained appropriately.Patients had been bladder scanned and found to have urine in their bladders or exchange occurs and urine flows.
 
Event or Problem Description
It was reported that there had since been 3 more events with the similar situation where the foley catheter was not drained appropriately.Patients had been bladder scanned and found to have urine in their bladders or exchange occurs and urine flows.Per customer follow up received via email on 04dec2025 stated that they did believe the device was discarded and 1 may have been picked up by clinical engineering.Patient retained urine and all devices had to be removed and replaced to allow urine to flow adequately.That was as much information they had at hand.They would need to go back into charts and review gender and ages.
 
Additional Manufacturer Narrative
The reported event was inconclusive because this investigation did not result in any additional findings and no sample was available for evaluation.The reported event is addressed within the labeling as it state.Visually inspect the product for any imperfections or surface deterioration prior to use.If package is opened or if any imperfection or surface deterioration is observed, do not use.Proper techniques for urinary catheter insertion.¿ perform hand hygiene immediately before and after insertion.¿ insert urinary catheters using aseptic technique and sterile equipment.¿ use the smallest foley catheter possible, consistent with good drainage.¿ document the indications for catheter insertion, date and time of catheter insertion, individual who inserted catheter, and date and time of catheter removal in patient record proper techniques for urinary catheter maintenance.¿ secure the foley catheter, use the statlock® foley stabilization device if provided.¿ maintain a closed drainage system by utilizing pre-connected, sealed catheter-tubing junctions.¿ maintain unobstructed urine flow and keep the catheter and collection tube free from kinking.¿ keep the collection bag below the level of the bladder or hips at all times ¿ empty the collection bag regularly (e.G., prior to transport) using a separate, clean collection container for each patient.¿ routine hygiene (e.G., cleansing of the meatal surface during daily bathing or showering) is appropriate.¿ leave foley catheter in place only as long as needed.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.No additional actions can be taken at this time.Udi format updated with available product information per gudid.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRICATH® TEMPERATURE SENSING FOLEY CATHETER
Common Device Name
TEMPERATURE SENSING FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key23701284
Report Number1018233-2025-10696
Device Sequence Number10628633
Product Code EZC
UDI-Device Identifier00801741046148
UDI-Public(01)00801741046148(11)250820(17)270930(10)NGKU0892
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K910846
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number119416M
Device Lot NumberNGKU0892
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/21/2025
Supplement Date Received by Manufacturer01/29/2026
Initial Report FDA Received Date12/02/2025
Supplement Report FDA Received Date02/05/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient SexUnknown
-
-