| Catalog Number |
NBD8 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/31/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2025, a patient underwent a procedure of percutaneous transhepatic cholangiography (ptcd) for obstructive jaundice caused by a pancreatic mass.On the third postoperative day, the drainage tube connector was ruptured, resulting in bile leakage.Despite bandaging, the leakage persisted.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: as the lot number for the device was provided, a review of the device history records was performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.Section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: manufacturing review: the device history records have been reviewed, and this lot met all release criteria.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.The investigation is inconclusive for the reported catheter hub fracture and fluid leak issues as no objective evidence was provided for review.A definitive root cause could not be determined based upon the provided information.A quality event has been opened to address the fracturing in navarre catheter hubs.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.G3, h6 (component) section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2025, a patient underwent a procedure of percutaneous transhepatic cholangiography (ptcd) for obstructive jaundice caused by a pancreatic mass.On the third postoperative day, the drainage tube connector was ruptured, resulting in bile leakage.Despite bandaging, the leakage persisted.There was no reported patient injury.
|
| |
|
Search Alerts/Recalls
|