MEDTRONIC PUERTO RICO OPERATIONS CO., HU CD HORIZON; APPLIANCE, FIXATION, SPINAL INTERLAMINAL Back to Search Results
Model Number 7540020
Medical Device Problem Code Device Slipped (1584)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/23/2025
Type of Reportable Event Malfunction
Event or Problem Description
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient having posterior lumbar spine therapy.It was reported that the clinician had a thread slip when screwing the pedicle screw into the set screw, which made it impossible to screw it in.Later, two new set screws were replaced during the operation.There were no patient symptoms reported.There were no further complications reported regarding the event.
 
Additional Manufacturer Narrative
H3: product analysis of product: (b)(4), lot: h6002694 visual and macroscopic examination revealed thread crest/flank damage; this damage appears to have initiated near the start of the thread and is evident around the damaged portion of the thread.Functional evaluation with a sample mas head found the set screw was unable to be fully engaged.This type of damage is consistent with misalignment of the mas head and set screw threads during construct assembly.E1: initial reporter is unknown.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO., HU
road 909, km. 0.4
humacao,pr
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO., HU
road 909, km. 0.4
humacao,pr
Manufacturer Contact
justin ellis
1800 pyramid place
memphis, TN 38132
7635265677
MDR Report Key23702973
Report Number2647346-2025-00231
Device Sequence Number16100845
Product Code KWP
UDI-Device Identifier00885074457093
UDI-Public(01)00885074457093(11)241025(10)H6002694
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K052187
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7540020
Device Catalogue Number7540020
Device Lot NumberH6002694
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/13/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/13/2025
Initial Report FDA Received Date12/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/25/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-