| Model Number |
72404133 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
Urinary Incontinence (4572)
|
| Date of Event |
11/06/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant product udi for balloon is (b)(4).Concomitant product udi for pump is (b)(4).The device is not available for analysis; therefore, no physical or visual analysis of the product could be performed.There was no report of a device performance allegation.The reported patient symptom of incontinence is a known risk associated with this device type and indicated as such in the instructions for use.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient with this artificial urinary sphincter (aus) was experiencing recurring incontinence.Clinical evaluation showed that the device was cycling appropriately, and it was determined that a smaller cuff was needed to improve continence.A surgical procedure was performed in which the existing cuff was removed and replaced with a smaller one, while the pressure-regulating balloon and control pump were retained.There were no further patient complications reported.
|
| |
|