| Catalog Number |
4251621-02 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
11/05/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
As reported by the user facility: reason of complaint: broke off in patient, potential infection risk.3.5cm of the iv catheter was retained in the patient's skin upon removal of the device.This patient then had to be transported to a higher acuity hospital to have the remaining catheter surgically removed.
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow-up will be submitted when the investigation results become available.
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).A review of the device history record (dhr) was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.No sample was provided for evaluation.Based on the data from the investigation we are unable to determine the root cause of the reported incident.The reported defect was unable to be confirmed.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|
| |
|
Search Alerts/Recalls
|