| Model Number |
NM-400U-0423 |
| Medical Device Problem Code |
Corroded (1131)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/03/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the single use injector had a rusty needle.This was found during preparation for use; there was no patient involved.
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported that the single use injector had a rusty needle.This was found during preparation for use; there was no patient involved.
|
| |
|
Additional Manufacturer Narrative
|
|
The mdr supplemental report is being submitted to provide updated fields and final investigation results.Updated fields: d8, h2, h3, h4, h6, h11 the device was returned for evaluation and the customer's allegation was not confirmed.A definitive root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: no problem detected.Olympus will continue to monitor the performance of this device.
|
| |
|
Event or Problem Description
|
|
No additional information received from the customer.
|
| |
|