|
It was reported that a patient underwent an atrial fibrillation ablation procedure with a thermocool smarttouch sf catheter and post procedure the bwi product analysis lab identified a hole on the surface of the pebax.During the procedure, after the map was made with the octaray in the left atrium, ablation was started.However, the visualization of ablation catheter jumped on carto.No actual movement of catheter in the left atrium was seen on fluoroscopy.The ablation catheter cable but the issue persisted.The catheter was exchanged and that resolved the issue.The case continued and was completed.No patient consequences were reported.
|
|
The product investigation was completed.Device evaluation details: the device was returned to johnson & johnson medtech (j&j medtech) for evaluation.Following j&j medtech procedures, a visual inspection and magnetic test of the returned device were performed.Visual inspection reddish material and a hole in the surface of the pebax.The device was connected to the carto 3 system and it was recognized and visualized correctly.No magnetic issues or jumping icon was observed.A manufacturing record evaluation was performed for the finished device lot number 31613947l and no internal action was found during the review.The reddish material inside the pebax could be related to the jumping icon reported by the customer; therefore, the customer complaint was confirmed.The potential cause of the pebax damage could be related to the manipulation of the device during the procedure; however, this cannot be conclusively determined.The instructions for use (ifu) contain the following recommendations: when cleaning the tip electrode, be careful not to twist the tip electrode with respect to the catheter shaft; twisting may damage the tip electrode bond and loosen the tip electrode, or may damage the contact force sensor.A significant change in the baseline reading after cleaning might indicate a damaged contact force sensor.In addition, in order to prevent damage to the catheter tip, verify compatibility between the sheath and catheter, advance the catheter through the sheath prior to insertion.Any sheath < 8.5 f is contraindicated.In relation to the jumping icon, the carto® 3 system instructions for use (ifu) contain the following information: ensure that catheter visualization initialization has finished.Ensure that the electrodes selected for mapping are out of sheath.Ensure that conditions for catheter visualization are fulfilled and the selected mapping catheter is visualized.Disconnect and reconnect the catheter and ensure proper connection.Replace the catheter or the cable.As part of johnson & johnson medtech¿s quality process, all devices are manufactured, inspected, and released to approved specifications.This product complaint will be addressed through johnson & johnson medtech's quality system.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|