| Model Number |
B31000 |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/31/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional review indicates this event was already reported in manufacturer report (#2182207-2025-02992).Any additional information regarding this event will be submitted as a supplemental submission to this report.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that in both cases, the surgeon could not lock the burr hole clip.It was unknown if there were any contributing factors.The surgeon tried to clean up the burr hole from bone segments and tried again but without success.The clip was replaced by another one.The issue was resolved.Additional information was received from a manufacturer representative (rep) reporting that the cases occurred in the same case, same day.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: analysis of the device had shown that the locking tab of the movable jaw on the returned clip failed to engage with the edge of the cam disk when the support clip was at the closed position.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|