• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION CORPORATION INFINION? PRO; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION CORPORATION INFINION? PRO; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF Back to Search Results
Model Number SC-2318-50
Medical Device Problem Codes Unexpected Therapeutic Results (1631); Migration (4003)
Health Effect - Clinical Code Pain (1994)
Date of Event 11/12/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the spinal cord stimulator (scs) leads of the patient had migrated out of the epidural space which caused the patient to experience loss of stimulation.The patient underwent a scs revision procedure, was doing well postoperatively and the explanted devices were kept by the facility.
 
Additional Manufacturer Narrative
Additional suspect medical device component involved in the event: product family: scs-linear leads.Upn: m365sc2318500.Model: sc-2318-50.Serial: (b)(6).Batch: (b)(6).Udi: (b)(4).Product family: scs-ipg-pc.Upn: m365sc14320.Model: sc-1432.Serial: (b)(6).Batch: (b)(6).Udi: (b)(4).Block d2b: qrb.
 
Additional Manufacturer Narrative
Model number/catalog number: sc-2318-50/ sc-1432.Serial number: (b)(6).Batch/lot number: 5005160/ 5005139/ 238351.Investigation results: the device was not returned for analysis; therefore, a technical analysis could not be performed.(good faith effort attempts were made to try and retrieve the device, however response was not received.).Device history record: it was confirmed this device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Labeling review : review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Investigation conclusion: based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of known inherent risk of device.
 
Event or Problem Description
It was reported that the spinal cord stimulator (scs) leads of the patient had migrated out of the epidural space which caused the patient to experience loss of stimulation.The patient underwent a scs revision procedure, was doing well postoperatively and the explanted devices were kept by the facility.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINION? PRO
Common Device Name
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION CORPORATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
erik sherburne
25155 rye canyon loop
valencia, CA 91355
7632920920
MDR Report Key23708299
Report Number3006630150-2025-11142
Device Sequence Number21500999
Product Code LGW
UDI-Device Identifier00191506018726
UDI-Public(01)00191506018726(17)270722(10)5005160
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
P030017
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberSC-2318-50
Device Catalogue NumberSC-2318-50
Device Lot Number5005160
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/12/2025
Supplement Date Received by Manufacturer03/09/2026
Initial Report FDA Received Date12/03/2025
Supplement Report FDA Received Date04/01/2026
Date Device Manufactured07/22/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
Patient SexFemale
-
-