| Model Number |
SC-2318-50 |
| Medical Device Problem Codes |
Unexpected Therapeutic Results (1631); Migration (4003)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
11/12/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the spinal cord stimulator (scs) leads of the patient had migrated out of the epidural space which caused the patient to experience loss of stimulation.The patient underwent a scs revision procedure, was doing well postoperatively and the explanted devices were kept by the facility.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional suspect medical device component involved in the event: product family: scs-linear leads.Upn: m365sc2318500.Model: sc-2318-50.Serial: (b)(6).Batch: (b)(6).Udi: (b)(4).Product family: scs-ipg-pc.Upn: m365sc14320.Model: sc-1432.Serial: (b)(6).Batch: (b)(6).Udi: (b)(4).Block d2b: qrb.
|
| |
|
Additional Manufacturer Narrative
|
|
Model number/catalog number: sc-2318-50/ sc-1432.Serial number: (b)(6).Batch/lot number: 5005160/ 5005139/ 238351.Investigation results: the device was not returned for analysis; therefore, a technical analysis could not be performed.(good faith effort attempts were made to try and retrieve the device, however response was not received.).Device history record: it was confirmed this device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Labeling review : review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Investigation conclusion: based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of known inherent risk of device.
|
| |
|
Event or Problem Description
|
|
It was reported that the spinal cord stimulator (scs) leads of the patient had migrated out of the epidural space which caused the patient to experience loss of stimulation.The patient underwent a scs revision procedure, was doing well postoperatively and the explanted devices were kept by the facility.
|
| |
|