MAUDE Adverse Event Report: C.R. BARD INC. (COVINGTON) -1018233 BARDEX® LUBRI-SIL® I.C. TEMPERATURE-SENSING ALL-SILICONE FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD INC. (COVINGTON) -1018233 BARDEX® LUBRI-SIL® I.C. TEMPERATURE-SENSING ALL-SILICONE FOLEY CATHETER Back to Search Results
Catalog Number 119316
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/21/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The initial reporter's zip code:(b)(6).The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The reported product number is unknown.The udi number for this product is not available.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the foley catheter balloon was deflated and they thought that popped.
 
Additional Manufacturer Narrative
The initial reporter's zip code: (b)(4).The reported event is unconfirmed.Visual: received one (1) used all silicone foley catheter attached to the urine meter bag without original packaging.Verified material number 119316 and manufacturing lot number ngkv2689.Visual inspection noted no obvious observations.The catheter balloon was inflated with 10ml methylene blue solution (3 drops 1percent aq methylene blue per 100ml distilled water) and the catheter rested with no leaks noted.The balloon passively deflated in under one (1) minute with no cuffing or leaks noted, returning 10ml of solution.This is within specification per inspection procedure which states, catheter leaks, valve leaks, balloon perforation, lumen to lumen, prematurely deflated and dislodged balloon are not allowed, and must inflate and deflate with the use of a syringe.As the reported event is unconfirmed, a risk and labeling review is not required.A review of the device history record was not necessary due to the reported event being unconfirmed.Based on the investigation results no additional action is required at this time.Corrections: d, f, g, h.Udi format updated with available product information per gudid.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the foley catheter balloon was deflated and they thought that popped.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRI-SIL® I.C. TEMPERATURE-SENSING ALL-SILICONE FOLEY CATHETER
Common Device Name
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key23708514
Report Number1018233-2025-10726
Device Sequence Number13424361
Product Code MJC
UDI-Device Identifier00801741039850
UDI-Public(01)00801741039850(11)250914(17)300827(10)NGKV2689
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K070582
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/03/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number119316
Device Lot NumberNGKV2689
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/24/2025
Supplement Date Received by Manufacturer04/03/2026
Initial Report FDA Received Date12/03/2025
Supplement Report FDA Received Date04/06/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/14/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-