The initial reporter's zip code:(b)(6).The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The reported product number is unknown.The udi number for this product is not available.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
The initial reporter's zip code: (b)(4).The reported event is unconfirmed.Visual: received one (1) used all silicone foley catheter attached to the urine meter bag without original packaging.Verified material number 119316 and manufacturing lot number ngkv2689.Visual inspection noted no obvious observations.The catheter balloon was inflated with 10ml methylene blue solution (3 drops 1percent aq methylene blue per 100ml distilled water) and the catheter rested with no leaks noted.The balloon passively deflated in under one (1) minute with no cuffing or leaks noted, returning 10ml of solution.This is within specification per inspection procedure which states, catheter leaks, valve leaks, balloon perforation, lumen to lumen, prematurely deflated and dislodged balloon are not allowed, and must inflate and deflate with the use of a syringe.As the reported event is unconfirmed, a risk and labeling review is not required.A review of the device history record was not necessary due to the reported event being unconfirmed.Based on the investigation results no additional action is required at this time.Corrections: d, f, g, h.Udi format updated with available product information per gudid.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|