It was reported that during a da vinci-assisted appendectomy procedure, loose staples from two sureform 30 reloads fell inside the patient.The staples were cleaned up, and no patient injury was reported.The customer opened a third stapler reload and it was normal with no loose staples.Intuitive surgical, inc.(isi) followed up with the initial reporter and obtained the following additional information: the sureform 30 reload was inspected prior to use and no damage was noted.According to the customer, the sureform 30 stapler instrument did not fire at all.Surgeon did not notice any issues with functionality of stapler reload.All of loose staples were retrieved and confirmed visually.There was no bleeding observed along staple line and there were no holes or gaps within staple line.Procedure was completed robotically with no injury to patient.No additional surgical procedures were required to remove the loose staples.Furthermore, no postoperative tests, such as ultrasound, were performed.Patient has not returned to hospital for any post-surgical complications.
|