STRYKER ORTHOPAEDICS-MAHWAH TRIDENT 0 DEG X3 INSERT 32E; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 723-00-32E
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 11/10/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.The following devices were also listed in this report: tridentii tritanium cluster54e; cat# 702-04-54e; lot# 98553101a, delta v-40 ceramic head 32/0; cat# 6570-0-132; lot# 12394951, standard insignia collared hip stem; cat# 7000-5505; lot# 18382752.It cannot be determined which, if any of these devices may have caused or contributed to the patient's experience.
 
Event or Problem Description
Postoperative infection following tha.Trident 2 and insignia removed on (b)(6) 2025.
 
Additional Manufacturer Narrative
Corrected data: manufacturing date and expiration date.An event regarding infection involving a trident liner was reported.The event was not confirmed.Method & results: -product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.-clinician review: no medical records were received for review with a clinical consultant.-product history review: review of the product history records indicate the reported device was manufactured with no reported relevant discrepancies.-complaint history review: there have been no other similar events for the lot referenced.There have been 2 other similar events for the sterile lot referenced, also related to infection.A review of both the sterilization data and microbiology data was performed for this sterile lot commonality.It was identified that the sterilization data was within specification and no microbiology excursions were noted for this lot and as such no further review is required.Conclusions: it was reported that the patient was revised due to infection.All stryker products are sold as sterile either by being validated to a minimum sterility assurance level sal of 10^-6 or aseptically filled under a validated process in accordance with applicable external standards.The exact cause of the event could not be determined because insufficient information was provided.Further information such as pathology reports, pre- and post-operative x-rays and the revision operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
No new information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT 0 DEG X3 INSERT 32E
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
brad curtis
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key23709219
Report Number0002249697-2025-01408
Device Sequence Number10649630
Product Code LPH
UDI-Device Identifier07613327342499
UDI-Public07613327342499
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K182468
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number723-00-32E
Device Lot NumberE163DY
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/10/2025
Supplement Date Received by Manufacturer03/27/2026
Initial Report FDA Received Date12/03/2025
Supplement Report FDA Received Date04/09/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/06/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-