• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. SIMPLYGO MINI; GENERATOR, OXYGEN, PORTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. SIMPLYGO MINI; GENERATOR, OXYGEN, PORTABLE Back to Search Results
Model Number 1113604
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/24/2025
Type of Reportable Event Malfunction
Event or Problem Description
The simplygo mini was returned to the manufacturer service center for evaluation without an initial customer complaint.During the evaluation it was noted that the main printed circuit assembly had thermal damage.Additionally, a tech fault was reported, the rear cover was found cracked, the o2 side assembly and airside assembly were found contaminated therefore they were replaced.In conclusion, the device's printed circuit assembly was replaced to address the issue.The device was tested and passed all final testing.A final report is being submitted.If new information becomes available, a follow up/supplemental report will be completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIMPLYGO MINI
Common Device Name
GENERATOR, OXYGEN, PORTABLE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key23709293
Report Number2518422-2025-112244
Device Sequence Number16132547
Product Code CAW
UDI-Device Identifier00606959032422
UDI-Public(01)00606959032422
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K111885
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 12/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number1113604
Device Catalogue Number1113604
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/24/2025
Initial Report FDA Received Date12/03/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/24/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-