• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GREATBATCH MEDICAL MYOPORE® SUTURELESS MYOCARDIAL PACING LEAD; BIPOLAR EPICARDIAL LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GREATBATCH MEDICAL MYOPORE® SUTURELESS MYOCARDIAL PACING LEAD; BIPOLAR EPICARDIAL LEAD Back to Search Results
Model Number 511212
Medical Device Problem Codes Failure to Capture (1081); Low impedance (2285); High Capture Threshold (3266)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/24/2025
Type of Reportable Event Malfunction
Event or Problem Description
Lead was implanted and had great results for about 5 days.Lead was firmly still in the epicardium and had started to endothelise.The entire epicardium was covered in significant scar and it was suspected that the lead was screwed into scar tissue.Lead was surgically abandoned and replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MYOPORE® SUTURELESS MYOCARDIAL PACING LEAD
Common Device Name
BIPOLAR EPICARDIAL LEAD
Manufacturer (Section D)
GREATBATCH MEDICAL
2300 berkshire lane
plymouth MN 55441
Manufacturer (Section G)
GREATBATCH MEDICAL
2300 berkshire lane
plymouth MN 55441
Manufacturer Contact
jacob lee
2300 berkshire lane
plymouth, MN 55441
MDR Report Key23710056
Report Number2183787-2025-00057
Device Sequence Number13443460
Product Code DTB
UDI-Device Identifier00821329900228
UDI-Public(01)00821329900228(11)241001(17)271001(20)00
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P130012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 12/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number511212
Device Lot NumberW9525254
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/10/2025
Initial Report FDA Received Date12/03/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/01/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;