| Model Number |
G141 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Failure to Capture (1081); High impedance (1291); Over-Sensing (1438); Pacing Problem (1439)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/19/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) recorded an alert for high out of range shock impedance measurements above 200 ohms on the non-boston scientific right ventricular (rv) lead.Technical services (ts) was consulted, and upon review it was noted that there was noise on the rv lead and non-boston scientific left ventricular (lv) lead, causing oversensing and pacing inhibition on the lv channel.Possible loss of capture (loc) and a suspected fracture was also noted on the rv channel.Ts advised on further in office evaluation of the system.As a result, this crt-d system was reprogrammed with tachy therapies off and to exclude the rv lead.No additional adverse patient effects were reported.This device remains in service.
|
| |
|
Additional Manufacturer Narrative
|
|
If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|