| Model Number |
511B |
| Medical Device Problem Code |
Gas Output Problem (1266)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/03/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during a surgical procedure involving the affinity nt oxygenator, there was insufficient oxygenation capacity and insufficient carbon dioxide removal capacity.The issue was identified as an alleged product performance problem with the oxygenator.Pco2 (partial pressure of carbon dioxide) was between 55.3mmhg and 55.8mmhg.Po2 (partial pressure of oxygen) was 115.3mmhg.Hct (hematocrit) was 32.7%.The device was replaced to complete the procedure.No patient complications have been reported as a result of this event.
|
| |
|