• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. DS2ADV AUTO CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. DS2ADV AUTO CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number FRX521H14C
Medical Device Problem Code Thermal Decomposition of Device (1071)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/19/2025
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information alleging an issue related to a ds2adv auto cpap device involving a thermal product malfunction.There was no report of patient harm or injury, and there was no report of medical intervention.The device was returned to the third-party service center, and the service center found a short circuit on the pca, burned components, and a burned smell inside.The ui panel flat cable was burned, the iso port was contaminated and had burn marks on it, and the inlet/outlet seal was contaminated and also had burn marks.The heater plate, blower box, blower and blower outlet seal, top enclosure, bottom enclosure, and center enclosure were all contaminated with particles and smelled burned inside.Due to the alleged malfunction, the device was forwarded to the manufacturer¿s product investigation lab (pil) for additional testing and investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DS2ADV AUTO CPAP
Common Device Name
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key23716487
Report Number2518422-2025-112231
Device Sequence Number18354329
Product Code BZD
UDI-Device Identifier00606959062757
UDI-Public(01)00606959062757
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K200480
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation 505
Type of Report Initial
Report Date (Section B) 12/04/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberFRX521H14C
Device Catalogue NumberFRX521H14C
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/27/2025
Initial Report FDA Received Date12/04/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/23/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-