ASCENSION ORTHOPEDICS, INC. PIP SZ. 30 PROXIMAL; HIGH DEMAND, REVISION, SEMI-CONSTRAINED, PYROLYTIC CARBON, UNCEMENTED FINGER PRO
|
Back to Search Results |
|
| Catalog Number |
PIP-200-30P-WW |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/06/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during a pip joint arthroplasty surgery, surgeon implanted the pip sz.30 proximal.Upon x-raying, there was a notable line through the implant.The implant was removed, and it came out in two pieces.All pieces were removed and accounted.The first piece was removed by hand and the remaining piece with a bur (disintegrated the implant and x-rayed to confirm all was removed) before implanting second implant.A new implant was then opened and used instead.The procedure was resumed, using an equivalent s+n back-up.No further complications were reported.It is unknown if a delay was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).This complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known risks with the device itself or with its use that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Search Alerts/Recalls
|
|
|