• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TMJ SOLUTIONS, LLC DBA TMJ CONCEPTS TMJ UNILATERAL IMPLANT; JOINT, TEMPOROMANDIBULAR, IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TMJ SOLUTIONS, LLC DBA TMJ CONCEPTS TMJ UNILATERAL IMPLANT; JOINT, TEMPOROMANDIBULAR, IMPLANT Back to Search Results
Catalog Number CHG010
Medical Device Problem Code Nonstandard Device (1420)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/16/2024
Type of Reportable Event Malfunction
Event or Problem Description
Within the scope evaluation for a product field action, this case was identified to have the potential to exhibit the hazard of the recommended screw length being greater than the bone thickness for implants with screw holes located in an area of thin bone.No adverse consequences were reported for this event.
 
Additional Manufacturer Narrative
Device is not available for evaluation.If additional information is received, it will be reported on a supplemental report.
 
Additional Manufacturer Narrative
Additional information: h4, h6.Correction: d2a.The reported event could not be confirmed.Stryker requested the customer to acknowledge the risks in relation to this product field action (pfa) (mentioned in the customer notification letter).For this case, the surgeon confirmed that there were no adverse consequences observed.
 
Event or Problem Description
No new information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TMJ UNILATERAL IMPLANT
Common Device Name
JOINT, TEMPOROMANDIBULAR, IMPLANT
Manufacturer (Section D)
TMJ SOLUTIONS, LLC DBA TMJ CONCEPTS
6059 king drive
ventura 93003
Manufacturer (Section G)
TMJ SOLUTIONS, LLC DBA TMJ CONCEPTS
6059 king drive
ventura 93003
Manufacturer Contact
elena drewenskus
boetzingerstr. 41
freiburg D-791-11
GM   D-79111
7614888208
MDR Report Key23716796
Report Number0002031049-2025-00106
Device Sequence Number8982184
Product Code LZD
UDI-Device Identifier07613327626551
UDI-Public07613327626551
Combination Product (Y/N)N
Initial Reporter CountryGM
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberCHG010
Device Lot Number2409241016
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/26/2025
Supplement Date Received by Manufacturer11/26/2025
Initial Report FDA Received Date12/04/2025
Supplement Report FDA Received Date04/24/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/10/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-