| Catalog Number |
960784 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
Limb Fracture (4518); Bone Notching (4890)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported there was a fracture in the left knee.A revision procedure took place on (b)(6) 2025.It was noted the bone behind the femoral component was in a very bad condition, there was no bony support for the metal component.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information that has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint recon described in this report.H11 additional narrative: added: a1, a3a, h6.Health effect - clinical code.If information that was not available for the initial report is obtained, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: d10 (concomitant).Corrected: h4.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4) investigation summary according to the information received, hcp is reporing a fracture in the left knee-tep.See x-rays.Revision is planned".The product was not returned to depuy synthes; however, an x-ray and photos were provided for review.(b)(4) implant passport and (b)(4) x-rays.Upon visual evaluation of the x-ray provided, it was possible to identify that the distal section of the sig fem adap +2/-2 offset bolt was fractured, it could be observed that the fractured piece appears to locate at the posterior knee section.The mode of failure of the device is multi-factorial and consideration must be given to all other potential influences such as surgical process, patient variables (e.G.Activity level and use), anatomical considerations and patient changes over time.Any conclusions from the investigational input provided have to be placed into context with all other relevant factors.A manufacturing record evaluation was performed for the finished device product code 960784, lot number 206081, and no non-conformances / manufacturing irregularities were identified during manufacturing.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was confirmed as the observed condition of the sig fem adap +2/-2 offset bolt would have contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot 1) quantity manufactured: (b)(4) pcs total manufactured 2) date of manufacture: 02-01-2012 3) any anomalies or deviations identified in dhr: an internal nr was found, however is not related with the reported condition.4) expiry date: 02-01-2022 5) ifu reference: ifu-0902-00-252.Device history review a manufacturing record evaluation was performed for the finished device product code 960784 lot number 206081, and no non-conformances / manufacturing irregularities were identified during manufacturing.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: d9.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|