ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) PROTEGE IPG, 16-CHANNEL RECHARGEABLE; SCS IPG Back to Search Results
Model Number 3789
Medical Device Problem Code Wireless Communication Problem (3283)
Health Effect - Clinical Code Inadequate Pain Relief (2388)
Date of Event 11/11/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported the patient¿s ipg was unable to communicate with external devices.A manufacturer representative confirmed the issue and the ipg deemed inoperable.Surgical intervention may be taken at a later date to address the issue.
 
Additional Manufacturer Narrative
Date of event is estimated.
 
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported event is consistent with the battery reaching end of life.
 
Event or Problem Description
It was reported patient underwent surgical intervention on (b)(6) 2025 wherein the ipg was explanted and replaced to address the issue.Therapy was restored post-op.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROTEGE IPG, 16-CHANNEL RECHARGEABLE
Common Device Name
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer Contact
sara speyerer
6901 preston road
plano, TX 75024
9723098000
MDR Report Key23719576
Report Number3006705815-2025-19921
Device Sequence Number18329186
Product Code LGW
UDI-Device Identifier05414734405690
UDI-Public(01)05414734405690(10)5215234(17)170831
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P010032
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2017
Device Model Number3789
Device Catalogue Number3789
Device Lot Number5215234
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/11/2025
Supplement Date Received by Manufacturer12/29/2025
Initial Report FDA Received Date12/04/2025
Supplement Report FDA Received Date01/20/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/01/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SCS LEAD.
Outcome Attributed to Adverse Event Other;
Patient Age52 YR
Patient SexFemale
Patient Weight77 KG
-
-