| Model Number |
3789 |
| Medical Device Problem Code |
Wireless Communication Problem (3283)
|
| Health Effect - Clinical Code |
Inadequate Pain Relief (2388)
|
| Date of Event |
11/11/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported the patient¿s ipg was unable to communicate with external devices.A manufacturer representative confirmed the issue and the ipg deemed inoperable.Surgical intervention may be taken at a later date to address the issue.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of event is estimated.
|
| |
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported event is consistent with the battery reaching end of life.
|
| |
|
Event or Problem Description
|
|
It was reported patient underwent surgical intervention on (b)(6) 2025 wherein the ipg was explanted and replaced to address the issue.Therapy was restored post-op.
|
| |
|
Search Alerts/Recalls
|