| Model Number |
OPTI1860COM18 |
| Medical Device Problem Code |
Premature Separation (4045)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Balt usa reference#: (b)(4).04nov2025 - additional information was received from the issuer regarding this incident.Therefore an updated reportability evaluation was documented which determined this incident should be upgraded to reportable based on the additional information received.The returned device was inspected in our quality laboratory.During our analysis: we observed the implant coil was not included with the device return.We observed no sign of detachment cycle being ran.We observed no visual damage to the detachment zone with the pet jacket, lead wires and solder joints intact.We observed multiple minor kinks along the hypotube.We observed the microcatheter was not returned.We conducted destructive testing of the delivery pusher to reveal the detachment profile of the sr thread.We observed the distal tip of the sr thread is opaque in color - indicating the thread endured an overload in tensile stress.Root cause of the reported issue can likely be attributed to an overload of tensile stress exerted on the implant coil.Upon device return, we observed the delivery pusher was returned with minor kinks along the hypotube.Moreover, we noted the detachment zone showed no visual signs of a detachment cycle being ran and the implant coil was not returned.To further analyze the unit, the body coil was deconstructed to reveal the tip of the sr thread which showed that the thread was opaque in color, indicating the thread endured an overload in tensile stress.It is possible if the implant coil were to have become damaged during attempts to reposition the coil system, this could have led to the implant coil encountering excessive friction and ultimately lead to the implant coil becoming detached upon retracting the coil system.Without the return of the implant coil its exact condition is unknown, therefore further analysis of the implant coil could not be performed.Furthermore if the associated microcatheter were to have become damaged or temporarily ovalized during the procedure, this could have forced excessive friction on the implant coil and led to the reported issue.Without the return of the associated microcatheter, further analysis could not be performed.It is indicated in the lhr that the unit underwent 100% inspection which verifies that the implant was free of damage at the point of final inspection.Review of the lot history records did not reveal any in-process or lot-specific issue that could account for the observation.No additional complaint against lot number: f241001051 has been made for the same issue.Comprehensive analysis of this failure mode has remained subject to monitoring for any unacceptable increase in trend.The submission of this report or related information to the fda, and its release by fda, does not reflect a conclusion by the party submitting this report of the fda that the report or related information is an admission that the manufacturer, their employees, or the device caused or contributed to the reportable event.
|
| |
|
Event or Problem Description
|
|
It was reported that: "during embolization of the splenic artery aneurysm of patient mecs 1 with an 18 x 60 coil complex, the body part of the upper coil ruptured near the microcoil inside the microcatheter and was removed from the assembly using the rigid hydrophilic guidewire 035 x 260." update 04nov2025 - additional information received from the issuer: "did the procedure involve tortuous anatomy? regarding the anatomy, we cannot say that there was severe tortuosity.We characterized the anatomy of this case without any observations.Will the microcatheter be available for return? no.Can you confirm if the device was implanted? no, the device was not implanted and it was replaced.Did the implant's microcoil detach from the delivery wire without the user pressing the disconnect button? that's exactly what happened.If the implanted microcoil detached, where did this occur (inside the aneurysm, inside the microcatheter, etc.)? the detachment occurred inside the microcatheter.Could you provide a more detailed explanation of what "removed from the assembly using the 035 x 260 rigid hydrophilic guidewire" means? the hydrophilic guidewire was passed through the spring and, with rotational movements, the microcoil coiled around the hydrophilic delivery wire, thus allowing the removal of the assembly: hydrophilic delivery wire, microcoil, and supporting diagnostic microcatheter." incident report form submitted for this complaint indicates that no patient injury was sustained.
|
| |
|
Search Alerts/Recalls
|
|