| Catalog Number |
PLSL3516 |
| Medical Device Problem Codes |
Fracture (1260); Material Deformation (2976)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
11/07/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that "when removing the lisfranc plate, the heads of the screws break 3.5 mm.".
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed.A review of the device history is not possible because the lot number was not communicated.Should additional information become available, it will be provided in a supplemental report.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction to h3(device evaluated by mfg) and h6(device code).The reported event could not be confirmed since the device was returned for evaluation and was observed to be deformed, not broken.The returned screw was visually inspected, revealing notable deformation of the torx recess on the screw head, which appeared completely distorted due to the application of excessive force.Prominent scratch marks were observed on the screw head, along with a shiny wear surface beneath the head.These marks indicate that the sides of the screw contacted the implant plate, suggesting that the screw was inserted under misalignment.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on investigation, the root cause was attributed to be user related.If more information is provided, the case will be reassessed.
|
| |
|
Event or Problem Description
|
|
It was reported that "when removing the lisfranc plate, the heads of the screws break 3.5 mm.".
|
| |
|