• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH LOCKING SCREW ANCHORAGE Ø3.5MM / L14MM; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH LOCKING SCREW ANCHORAGE Ø3.5MM / L14MM; PLATE, FIXATION, BONE Back to Search Results
Catalog Number PLSL3514
Medical Device Problem Codes Fracture (1260); Material Deformation (2976)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 11/07/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "when removing the lisfranc plate, the heads of the screws break 3.5 mm.".
 
Additional Manufacturer Narrative
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed.A review of the device history is not possible because the lot number was not communicated.Should additional information become available, it will be provided in a supplemental report.
 
Additional Manufacturer Narrative
Correction to h3(device evaluated by mfg) and h6(device code).The reported event could not be confirmed since the device was returned for evaluation and was observed to be deformed, not broken.The returned screw was visually inspected and showed distinct deformation marks around the screw head.At the torx recess, markings were observed along the edges, indicating that the screwdriver slipped during use.Additionally, a small dent was noted on one side of the screw head.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on investigation, the root cause was attributed to be user related.If more information is provided, the case will be reassessed.
 
Event or Problem Description
It was reported that "when removing the lisfranc plate, the heads of the screws break 3.5 mm.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCKING SCREW ANCHORAGE Ø3.5MM / L14MM
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ   2545
Manufacturer Contact
dannielle stap
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key23720331
Report Number0008031020-2025-02131
Device Sequence Number8835586
Product Code HRS
UDI-Device Identifier07613252262343
UDI-Public(01)07613252262343(10)AM5470
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K083447
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberPLSL3514
Device Lot NumberAM5470
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/07/2025
Supplement Date Received by Manufacturer01/22/2026
Initial Report FDA Received Date12/04/2025
Supplement Report FDA Received Date02/17/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/07/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-