MEDOS INTERNATIONAL SÃ RL HEALIX AV BR 4.5 TAPE WHT/BLUE; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
| Catalog Number |
222360 |
| Medical Device Problem Codes |
Break (1069); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/19/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a arthroscopic rotator cuff repair surgery, after creating a bone foramen with an awl, the surgeon attempted to insert a healix av br 4.5 tape wht/blue, but the anchor would not advance and the device broke.The fragments were retrieved.Another same product was used without any problems.The surgeon attempted to thread the specified number of sutures through the 4.9mm healx adv sp bioc anchor but felt strong resistance.After somehow managing to insert the thread into the anchor, the surgeon operated it as usual when he found that the anchor had cracked, so it would not be used.Another product was used, and the surgery was completed successfully.There was no surgical delay and no harm to the patient.The device was brand new and the first use when the issue occurred.Report 1 of 2 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: the complaint device was discarded by the customer, therefore not returned to depuy synthes and unavailable for a physical evaluation.A manufacturing record evaluation was performed for the finished device lot number (265l126), and no non-conformance was identified.Based on the current available information, we cannot determine a root cause for the reported failure.If additional information is received in the future, we will reopen the complaint and perform the investigation as appropriate.It has been determined that no corrective and/or preventative action is proposed and there is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
|
| |
|
Search Alerts/Recalls
|
|
|