MAUDE Adverse Event Report: SYNTHES GMBH TIBIAL NAIL-ADVANCED / 9MM 375MM / STERILE; NAIL, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH TIBIAL NAIL-ADVANCED / 9MM 375MM / STERILE; NAIL, FIXATION, BONE Back to Search Results
Catalog Number 04.043.145S
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Nonunion/Delayed-union Bone Fracture (2369)
Date of Event 11/15/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient underwent the tibial nail implant removal procedure due to breakage.Per surgeon the break was likely due to the +3 month non-union of a complex proximal tibial fracture.The broken nail was successfully removed and a new tibial nail with additional plating and bone graft was inserted into the patient.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: h4.H6: the product was not returned to depuy synthes; however, photos were provided for review.The photo investigation revealed the tibial nail advanced ø9 l 375 broken from the proximal tip.While no root cause can be established with the available information, factors such as the patient's age, underlying disease, bone density, or a possible early weight-bearing, could have led to implant failure.As the device was not returned, and as-received condition could not be assessed, a dimensional inspection and document/specification review were not completed.The overall complaint was confirmed as the observed condition of the tibial nail advanced ø9 l 375 would contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a dhr evaluation was performed for the finished device product code: 04.043.145s.Lot number: 24844p4.It was electronically reviewed and no nonconformances / manufacturing irregularities were identified during the manufacturing process.The product was released on: 12-july-2024.Manufacturing site: jabil bettlach.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL NAIL-ADVANCED / 9MM 375MM / STERILE
Common Device Name
NAIL, FIXATION, BONE
Manufacturer (Section D)
SYNTHES GMBH
luzernstrasse 19-21
zuchwil 4528
SZ  4528
Manufacturer (Section G)
WERK BETTLACH (CH)
muracherstrasse 3
bettlach 2544
SZ   2544
Manufacturer Contact
elba bello
1302 wrights lane east
west chester, PA 19380
9083863534
MDR Report Key23720711
Report Number8030965-2025-12187
Device Sequence Number10209473
Product Code JDS
UDI-Device Identifier10886982297655
UDI-Public(01)10886982297655(17)340531(10)24844P4
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K201336
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number04.043.145S
Device Lot Number24844P4
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/17/2025
Supplement Date Received by Manufacturer12/31/2025
Initial Report FDA Received Date12/04/2025
Supplement Report FDA Received Date01/06/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/12/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-