| Model Number |
RT385 |
| Medical Device Problem Code |
Disconnection (1171)
|
| Health Effect - Clinical Code |
Hypoxia (1918)
|
| Date of Event |
10/13/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Fisher & paykel (f&p) healthcare is currently in the process of finalizing the investigation.We will provide a follow up report upon completion of our investigation.
|
| |
|
Event or Problem Description
|
|
A healthcare facility in china reported that a ventilator alarmed for "disconnection of the tubing" of an rt385 adult ventilator circuit dual heated with evaqua technology.There was no patient harm reported.
|
| |
|
Event or Problem Description
|
|
A healthcare facility in china reported that, during patient use, a ventilator alarmed for "disconnection of the tubing" of an rt385 adult ventilator circuit dual heated with evaqua technology.There was no patient harm reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Method: the subject rt385 breathing circuit was not returned to fisher & paykel healthcare new zealand for evaluation.Our evaluation is therefore based on the photographs and the information provided by the healthcare facility, and our knowledge of the product.Results: the heatlhcare facility additionally reported that the device passed the ventilator pre-use check.Review of the provided information and photographs could not identify the cause of the reported event.Conclusion: based on the limited information provided by the healthcare facility and without the return of the subject device, we are unable to confirm the reported issue or determine the cause of the reported event.All rt385 adult dual-heated evaqua2 breathing circuit are visually inspected, as well as pressure and flow tested during production, and those that fail the test are rejected.The subject breathing circuit would have met the required specifications at the time of production.The user instructions that accompany the rt385 adult dual-heated evaqua2 breathing circuit show in pictorial format the correct set-up of the circuit and also states the following: "check all connections are tight before use." "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." "ensure appropriate ventilator or flow source alarms are set before connecting breathing set to patient.".
|
| |
|